3 Remote Interface Reference
118 E364xA User’s and Service Guide
– To ensure synchronization when the bus source is selected, send the *WAI
(wait) command. When the *WAI command is executed, the power supply
waits for all pending operations to complete before executing any additional
commands. For example, the following command string guarantees that the
first trigger is accepted and is executed before the second trigger is
recognized.
– You can use the *OPC? (operation complete query) command or the *OPC
(operation complete) command to signal when the operation is complete. The
*OPC?
command returns 1 to the output buffer when the operation is
complete. The *OPC command sets the OPC bit (bit 0) in the Standard Event
register when the operation is complete.
Immediate triggering
– To select the immediate trigger source, send the following command.
–When the IMMediate is selected as a trigger source, an INITiate command
immediately transfers the VOLT:TRIG or
CURR:TRIG value to VOLT or CURR
value. Any delay is ignored.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
TRIG:SOUR IMM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E364 A Series and is the answer not in the manual?

Keysight Technologies E364 A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE364 A Series
CategoryPower Supply
LanguageEnglish

Related product manuals