4 Error Messages
152 E364xA User’s and Service Guide
Overview
Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the
first error that was stored. Errors are cleared as you read them over the remote
interface. When you have read all errors from the queue, the ERROR annunciator
turns off and the errors are cleared. The power supply beeps once each time an
error is generated.
If more than 20 errors have occurred, the last error stored in the queue (the most
recent error) is replaced with -350, “Queue overflow”. No additional errors are
stored until you remove errors from the queue. If no errors have occurred when
you read the error queue, the supply responds with +0, “No error” over the remote
interface or “NO ERRORS” from the front panel.
The error queue is cleared by the *CLS (clear status) command or when power is
cycled. The errors are also cleared when you read the queue.
Front panel operation
Press (Local) key to return to front-panel operation mode if the power supply
is in remote operation mode.
ERRORS
If the ERROR annunciator is on, press key to view the errors. Use the knob to
scroll through the error numbers. Press to view the text of the error message.
Press or key to decrease or increase the scrolling speed of the text on the
display. All errors are cleared when you exit the menu by pressing the key or
let the display time out for about 30 seconds.
NOTE
The *RST (reset command) command does not clear the error queue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E364 A Series and is the answer not in the manual?

Keysight Technologies E364 A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE364 A Series
CategoryPower Supply
LanguageEnglish

Related product manuals