Error Messages 4
E364xA User’s and Service Guide 165
713
Bad readback cal data
The specified readback calibration values (CAL:VOLT or CAL:CURR) are out of
range. Note that the new calibration constants are not stored in the non-volatile
memory.
714
Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
717
Cal OVP status enabled
Overvoltage protection status is enabled. You must set overvoltage protection
status to OFF before and during the calibration.
718
Gain out of range for Gain Error Correction
The slope of the DAC gain is out of range. Hardware fails.
740
Cal checksum failed, secure state
741
Cal checksum failed, string data
743
Cal checksum failed, store/recall data in location 1
744
Cal checksum failed, store/recall data in location 2
745
Cal checksum failed, store/recall data in location 3
746
Cal checksum failed, DAC cal constants
747
Cal checksum failed, readback cal constants
748
Cal checksum failed, GPIB address
749
Cal checksum failed, internal data
754
Cal checksum failed, store/recall data in location 4
755
Cal checksum failed, store/recall data in location 5
Table 4-3 Calibration error messages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E364 A Series and is the answer not in the manual?

Keysight Technologies E364 A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE364 A Series
CategoryPower Supply
LanguageEnglish

Related product manuals