Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 253
Baseband Generators
A2 Vector BBG Assembly Troubleshooting
Figure 12-5 I and Q Input Connections – Front Panel
Rear Panel BNC Connectors
There are multiple rear panel BNC connectors on the A2 Vector BBG assembly.
For these to be functional one of the baseband generator options must be
installed in the instrument (653, 655, 656, 657). The I bar and Q bar outputs
also require that option 1EL be installed. The BB TRIG 1, BB TRIG 2, EVENT 1,
and PAT TRIG connectors are all bidirectional. To verify the functionality of
these connectors, follow the procedures below:
I OUT and Q OUT
To verify the functionality of the I and Q outputs use the following procedure,
which will require the use of an oscilloscope:
1. Return the instrument to a known state by pressing Preset.
2. Connect I OUT and Q OUT to channels 1 and 2 of an oscilloscope.
3. Setup the oscilloscope with the following settings:
— All Channels volts per division = 100 mV
— Horizontal scale per division = 500 ns

Table of Contents

Other manuals for Keysight Technologies N5166B CXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5166B CXG and is the answer not in the manual?

Keysight Technologies N5166B CXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5166B CXG
CategoryPortable Generator
LanguageEnglish

Related product manuals