Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 365
Assembly Replacement
A4 SSD (N5166B, N5172B, and N5182B only)
A4 SSD (N5166B, N5172B, and N5182B only)
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 15-17 and Figure 15-18 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Cover” on
page 336 and “Inner Bottom Cover” on page 340.
3. Place the signal generator flat and upside down with the rear panel facing
you.
4. Using the T-10 driver, remove the four screws (6) that attach the A4 SSD
assembly to the A3 RF assembly.
5. Remove the A4 SSD assembly by gently rocking it side to side to release
the pins.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Perform the post-repair procedures that pertain to this replacement
procedure. See Chapter 16, “Pre and Post-Repair Procedures.”

Table of Contents

Other manuals for Keysight Technologies N5166B CXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5166B CXG and is the answer not in the manual?

Keysight Technologies N5166B CXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5166B CXG
CategoryPortable Generator
LanguageEnglish

Related product manuals