Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 363
Assembly Replacement
A3BT1 Battery
A3BT1 Battery
Tools Required
— T-10 driver
— small flat-head screw driver
Removal Procedure
Refer to Figure 15-16 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Cover” on
page 336 and “Inner Bottom Cover” on page 340.
3. Place the signal generator flat with the A3 RF assembly facing up and the
front-panel towards you.
4. Using the flat-head screw driver, remove the A3BT1 by carefully prying the
battery out of its socket.
Replacement Procedure
1. Reverse the steps of the removal procedure.
2. Perform the post-repair procedures that pertain to this replacement
procedure. See Chapter 16, “Pre and Post-Repair Procedures.”
This battery contains lithium. Do not incinerate or puncture this battery.
Do not install this battery backwards. To dispose of the battery in a safe
manner. Refer to
“Lithium Battery Disposal” on page 35.
The battery is part of the power off circuitry. Removing the battery will not
cause any stored data to be lost.
To install the battery, the positive side is aligned with the positive sign on
the A3 RF assembly’s battery clip.