Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 417
Pre and Post-Repair Procedures
Post Repair Procedures
Table 16-3 Post Repair Procedures for N5173B, N5183B
Post-Repair Procedures
A1 Power Supply
A3 RF Assembly
A3BT1 Battery
A5 CPU Assembly
A6A1 Front Panel Interface
A6A3 Power Switch Assembly
A6A4 Keyboard Assembly
A6A5 Display LCD Assembly
A6MP1 Main Keypad
A7 Micro-Deck Assembly
Utilities
a
Calibration Array Restore X X X
Reset Mechanical Relays X
Additional Tasks
b
Install Instrument Firmware X X X
Restore Licenses X
Backup Licenses X
Set Time and Date X X X
Set Hostname X
Front Panel Tests XXXXX
Self Test
c
XXXXXXXXXX
Adjustments
a
RF ABUS Calibration X
MW ABUS Calibration X
MW Multiplier Prelevel Calibration X
MW Low Band Absolute Power Calibration X X
Op Amp Band Power Calibration X X
MW High Band Absolute Power Calibration X
MW ALC Calibration X
MW Open Loop Power Calibration X
AM Calibration (Option UNT) X X
Backup Cal Data to Motherboard Memory X X
Performance Verification
a
Self Test X X