EasyManuals Logo

Keysight Technologies N5172B EXG Service Guide

Keysight Technologies N5172B EXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #264 background imageLoading...
Page #264 background image
264 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Baseband Generators
A7 Real-Time BBG Assembly Troubleshooting
If all the power supply LEDs, as well as DS17, are on, proceed to the
next troubleshooting procedure.
If DS17 is off, replace the A7 Real-Time BBG assembly.
Figure 12-12 A7 Real-Time BBG Assembly Power Supply LED Locations
Self Test
There are a series of Self Tests for the A7 Real-Time BBG assembly. If the
previous troubleshooting procedures in the section have already been
performed and no fault was found, run the 1500 series Self Tests. For
information on running the instrument Self Tests see Chapter 4, “Self Test”
If any of the 1500 series Self Tests fail, and the previous troubleshooting
procedures found in this section have not found a fault, the most probable
cause of these failures is the A7 Real-Time BBG assembly.

Table of Contents

Other manuals for Keysight Technologies N5172B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5172B EXG and is the answer not in the manual?

Keysight Technologies N5172B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5172B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals