388 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
RF Connector
RF Connector
Tools Required
— T-10 driver
— 5/16” open-ended wrench
Removal Procedure
Refer to Figure 15-34 for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover from the signal generator. Refer to “Outer Cover”
on page 336.
3. Remove the front panel from the signal generator. Refer to “Front-Panel”
on page 342.
4. Using the 5/16” open-ended wrench, disconnect W9 (2) from the RF
connector.
5. Using the T-10 driver, remove the three screws (1) that attach the RF
connector to the chassis.
6. Remove the RF connector.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for the three screws is 9 in–lbs.
3. Recommended torque for W9 cable is 9 in–lbs.