418 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Pre and Post-Repair Procedures
Post Repair Procedures
Maximum Output Power X X
Power Level Accuracy X X
Harmonic Spurious X X
Sub-Harmonic Spurious X X
Non-Harmonic Spurious X X
AM Depth Accuracy (Option UNT) X X
AM Distortion (Option UNT) X X
FM Distortion (Option UNT) X X
FM Deviation Accuracy (Option UNT) X X
AM Frequency Response (External) (Option UNT) X X
Frequency Switching Speed (Option UNZ) X X
DCFM Carrier Offset (Option UNT) X X
Internal Pulse Modulation Level Accuracy (Option UNW) X X
Internal Pulse Modulation Rise/Fall Time (Option UNW) X X
Single-Sideband Phase Noise X X
a. For further information refer to Chapter 18, “Performance Verification and Adjustment Software.”
b. See the “Additional Tasks” section in this chapter for detailed instructions on these tasks.
c. For information on performing the instrument Self Tests refer to Chapter 4, “Self Test.”
Table 16-3 Post Repair Procedures for N5173B, N5183B
Post-Repair Procedures
A1 Power Supply
A3 RF Assembly
A3BT1 Battery
A5 CPU Assembly
A6A1 Front Panel Interface
A6A3 Power Switch Assembly
A6A4 Keyboard Assembly
A6A5 Display LCD Assembly
A6MP1 Main Keypad
A7 Micro-Deck Assembly