6 Repair Guide
110 Keysight N8262A Service Guide
Lift and remove calibrator cable assembly from
the mother board assembly at J8.
Step 1: Lift and remove the coaxial cable
labeled 1 (trig. in) from P1.
Step 2: Lift and remove the coaxial cable
labeled 2 (trig. out) from P2.
Note:
These coaxial cable are originally routed on the
right side of the chassis along the wall. See the
figure to the right for more details.