Commands GPIB VXI-11
Set the output voltage: VOLT<n> 0.24 ms 0.65 ms
Return the output setting: OUTP? 0.30 ms 1.35 ms
Set the unitto the reset state: *RST 5.01 ms 5.26 ms
10 point measurement
Return 10 point measurement: MEAS:VOLT? 3.00 ms 3.30 ms
Return 10 point fetch: FETC:VOLT? 0.49 1.49 ms
1 NPLC measurement
Return 1 NPLC measurement: MEAS:VOLT? 21.03 ms 20.96 ms
Return 1 NPLC fetch: FETC:VOLT? 0.63 ms 1.41 ms
25 k point measurement
Return 25 k point measurement: MEAS:VOLT? 521.0 ms 521.1 ms
Return 25 k point fetch: FETC:VOLT? 5.07 ms 7.01 ms
Return 25 k point ASCII array fetch: FETC:ARR:VOLT? 4009.5 ms 1010.8 ms
Return 25 k point binary array fetch: FETC:ARR:VOLT? 694.25 ms 30.39
5 SCPI Programming Reference
174 Keysight RP7900 Series Operating and Service Guide