Loading...
Parameter Typical Return
<block data> <block data>
Sends the block tate back to the instruemnt: SYST:SET <block data>
SYSTem:TIME <hh>, <mm>, <ss>
SYSTem:TIME?
Sets the time of the system clock. Specify hours (0 to 23), minutes (0 to 59), and seconds (0 to 59). The
real time clock is only used in conjunction with the Keysight 14585A Control and Analysis software.
Parameter Typical Return
<hh>,<mm>,<ss> <hh,mm,ss>
Set the clock to 8:30 PM: SYST:TIME2018,06,30
SYSTem:VERSion?
Returns the SCPI version that the instrument complies with. Cannot be determined from front panel.
Parameter Typical Return
(none) <"version">
Return the SCPI version: SYST:VERS?
l The command returns a string in the form "YYYY.V", where YYYY represents the year of the version
and V represents a version for that year.