Keysight Signal Generators Programming Guide 197
 Programming the Status Register System
Status Groups
Data Questionable Power Event Enable Register
The Data Questionable Power Event Enable Register lets you choose which bits in the Data 
Questionable Power Event Register set the summary bit (bit 3 of the Data Questionable Condition 
Register) to 1.
Command:
STATus:QUEStionable:POWer:ENABle <value> where <value> is the sum of the decimal values of 
the bits you want to enable
Example: Enable bit 3 and bit 2 so that whenever either of those bits are set to 1, the Data Questionable Power 
summary bit of the Data Questionable Condition Register is set to 1. Send the command 
STAT:QUES:POW:ENAB 520 (8 + 4). 
Query: STATus:QUEStionable:POWer:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the 
STATus:QUEStionable:POWer:ENABle <value> command.