Keysight E4980A/E4980AL Precision LCR Meter -129
Error Messages
Error Messages
H
70 Handler I/F not installed
The handler interface function is set to ON in a model that does not have a
handler interface board (option 201).
-114 Header Suffix out of Range
The header suffix is out of range.
I
73 I BIAS I/F not installed
The DC bias interface function is set to ON in a model that does not have the
DC bias interface board (option 002).
-224 Illegal parameter value
The parameter value is not properly set. This error occurs, for example, when
the :FUNC:IMP command is used to specify an Impedance-parameter that
does not exist (as CPRS).
51 Inconsistent limit setting
For the comparator function, user attempts to set the threshold in the
sequential mode by sending the COMP:SEQ:BIN command, although the
tolerance mode is now selected. Select the sequential mode using
COMP:MODE.
-161 Invalid block data
Invalid block data was received (for example, END received before length
satisfied).
-101 Invalid character
Invalid character was received.
-141 Invalid character data
Bad character data or unrecognized character data was received.
-121 Invalid character in number
Invalid character in numeric data.
-171 Invalid expression
Invalid expression was received (for example, illegal character in expression).
-103 Invalid separator
The message unit separator (for example, “;”, “,”) is improper.
-151 Invalid string data