20 Keysight E4980A/E4980AL Precision LCR Meter
General Information
Required Equipment
1-
Required Equipment
Table 1-1 lists the recommended equipment for performance test, adjustment
and troubleshooting for E4980A/ E4980AL.
Table 1-1 Recommended Test Equipment
Equipment Critical Specifications Recommended Model Qty.
Use
1
Frequency Counter Frequency Range: 1 MHz
Time Base Error < 50 ppm/year
Keysight 53181A 1 P
Digital Multimeter No Substitute Keysight 3458A 1 P,A
Standard Capacitors Set No Substitute 16380A 1 P,A
Standard Capacitors Set No Substitute 16380C with option
001
1P
Four Terminal Pair
Standard Resistor Set
No Substitute Keysight 42030A 1 P,A
4TP Open Termination No Substitute Keysight 42090A 1 P,A
4TP Short Termination No Substitute Keysight 42091A 1 P,A
Test Leads (BNC), 1 m No Substitute Keysight 16048A 1 P,A
Test Leads (BNC), 2 m No Substitute Keysight 16048D 1 P,A
Test Leads (BNC), 4 m No Substitute Keysight 16048E 1 P,A
DC Power Supply + 5V, 0.1A Keysight E3610A 1 T
Handler Simulator No Substitute p/n 04278-65001 1 T
Scanner Simulator No Substitute p/n 04278-65301 1 T
Bias Interface Simulator No Substitute p/n 42841-65001 1 T
Bias Interface Cable GP-IO Cable p/n 42841-61640 1 T
Interface Box No Substitute p/n 04294-61002 1 P,A
Cables BNC(m)-BNC(m) Cable, 122 cm p/n 8120-1840 1 P,A
Adapter BNC(f)-BNC(f) Adapter p/n 1250-1830 4 P,A
Dual Banana-BNC(f) Adapter p/n 1251-2277 1 P,A
1. P:Performance Test A:Adjustment T:Troubleshooting