Keysight E4980A/E4980AL Precision LCR Meter -131
Error Messages
Error Messages
P
-220 Parameter error
When a parameter-related error other than Errors -221 through -229 occur,
this error message is displayed.
-108 Parameter not allowed
The number of parameters exceeds that required for the command. For
instance, when a program message “:FORM:BORD NORM, ASC” is sent instead
of a correct program message with a command “:FORM:BORD NORM”, which
requires a parameter, the instrument receives the message that the number of
parameters is invalid. See the command reference to confirm the required
number of parameters.
1080 Power failed
Power unit hardware failure is detected. Contact Keysight Technology’s Sales
and Service Office or the company from which you bought the device.
1000 Power on test failed
An error occurs during a self test at power-on. The E4980A/ E4980AL may be
faulty. Contact Keysight Technology’s Sales and Service Office or the company
from which you bought the device.
A self test fails when an interface board is removed or not connected properly.
Check the connection with a cable when you change the setting of the
interface board.
-112 Program mnemonic too long
Program mnemonic is too long (maximum length is 12 characters).
Q
-350 Queue overflow
There is no room in the queue and an error occurred but was not recorded.
Clear the error queue.
-430 Query DEADLOCKED
Input buffer and output buffer are full; cannot continue.
-400 Query error
Query is improper.
-410 Query INTERRUPTED
Query is followed by DAB or GET before the response is completed.
-420 Query UNTERMINATED
Addressed to talk, incomplete program message received.