34 Keysight E4980A/E4980AL Precision LCR Meter
Troubleshooting
Power On Sequence Troubleshooting
4-
Figure 4-3 Output Voltage Location For MY46516001 (E4980A), SG46500501 (E4980A
GSA), MY54304302 (E4980AL), SG54300101 (E4980AL GSA) and above.
• If the voltmeter reading is not within the following limits, replace the power
supply unit.
Table 4-1 Power Supply Unit Output
Supply Range
+ 3.3 V + 3.2 to + 3.4
+ 5 V + 4.8 to + 5.2
+ 12 V + 11.4 to + 12.6