46 Keysight M8050A High-Performance BERT Getting Started Guide
3 Maintenance
Locating Electronic Manuals and Online Help
Various electronic manuals and the M8000 Series Online Help provide
information on how to configure and use the supported instrument
modules.
On installing the M8070B software, you will find documentation by clicking
Start > All Programs > Keysight M8070B > Keysight M8070B
Documentation.
For module driver related documents, navigate to:
C:\Program Files\Keysight\M8070B\Modules\
%ModuleDriverPackageName%\doc
where <ModuleDriverPackageName> is the name of the folder
corresponding to the package of interest.
You can also visit www.keysight.com/find/M8050A to find the latest
versions of related manuals and the data sheet for each M8050A module.
Routine Care
Except for performing initial chassis verification or troubleshooting, do
not operate the chassis with empty slots. Always insert a filler panel in
empty slots. This is especially important for the slots on either side of an
installed instrument module. This allows proper air flow and cooling, and
provides EMI shielding for the chassis and installed components. Leaving
slots empty can increase fan speed, raise ambient noise, overheat
components, and can cause the module to shut down.
Do not block the vent holes on the chassis. This overheats and damages
their components. Leave a gap of at least 2” (50 mm) around all vent
holes.