5 – Front Panel Operation 
66 
Set transient operation for Pulse mode  
In this example, assume that the CC mode is active, the slew rate is at the factory default setting (maximum 
rate), and the applicable transient operation parameters have been set as follows: pulse mode, main current level 
= 5 A, transient current level = 10 A, pulse width  = 1 millisecond. 
In this example, the electronic load starts conduction at the main current level setting (5 amps).  When the 
transient mode is turned on and an external trigger signal is received, module input current will slew to and 
remain at 10 A for the remainder of the 10 millisecond time period, then slew to and remain at 5 A until another 
trigger is received.  Any triggers that occur during the time the transient level is in effect will extend the pulse 
by another 10 milliseconds. Figure 5-6 shows the waveform that would result in this pulsed transient example. 
 
Figure 5-6.  Pulsed Transient Operation 
 
On the Function keypad, press Current.  On the Entry keypad, press 5 Enter . 
2.  On the Function keypad, press Current.  On the Function keypad, press  key until 
transient current is displayed. 
C:TLEV 
On the Entry keypad press  10 Enter. 
4.  On the Function keypad, press Tran.  On the Function keypad, press  key until 
transient pulse width is displayed. 
T:TWID 
On the Entry keypad press 0.01 Enter. 
6.  On the Function keypad, press Tran.  On the Function keypad, press  key until 
transient mode is displayed. 
T:MODE 
7. 
To select the pulse mode, press the 
Input or 
 Input key until the Pulse mode 
is displayed.  Then press Enter. 
T:MODE PULS 
8. 
On the Function keypad, press the Tran key and select ON to activate the transient 
TRAN ON 
9. 
On the Function keypad, press Trigger Control, then press Enter. This initiates 
INIT:IMMED 
Press Shift and Trigger. This starts the pulse. Repeat this step for additional pulses. 
Set transient operation for Toggled mode  
In this example, assume that the CC mode is active, the slew rate is at the factory default setting (maximum 
rate), and the applicable transient operation parameters have been set as follows: toggle mode, main current 
level = 5 A, transient current level = 10 A. 
The operation of toggled mode is similar to that described for continuous and pulse operation, except that each 
time a trigger is received the input alternates between the main and transient current levels. Figure 5-7 shows the 
waveform that would result for this toggled transient operation example