B - Performance Test and Calibration 
86 
Keysight N3304A Verification Test Record 
Constant Current Mode Tests 
60 Ampere Range Programming and Readback 
 Low Current  ( 1 A ) 
 Front Panel Display 
 High Current   ( 60 A ) 
 Front Panel Display 
6 Ampere Range Programming and Readback 
 Low Current  ( 1 A ) 
 Front Panel Display 
 High Current  ( 6 A ) 
 Front panel Display  
0.984 A 
Aout –10.5mA 
59.925 A 
Aout – 40mA 
 
0.9915 
Aout – 5.5mA 
5.9865 A 
Aout – 8mA 
_______A 
_______A 
_______A 
_______A 
 
_______A 
_______A 
_______A 
_______A 
1.016 A 
Aout + 10.5mA 
60.075 A 
Aout +40mA 
 
1.0085 
Aout + 5.5mA 
6.0135 A 
Aout + 8mA 
5 Amp Regulation at input voltage delta 4V to 60V 
 
60 Volt Range Programming and Readback 
 Low Voltage   ( 3 V ) 
 Front Panel Display 
 High Voltage  ( 60 V ) 
 Front Panel  Display 
6 Volt Range Programming and Readback 
 Low Voltage  ( 3 V ) 
 Front Panel Display 
 High Voltage  (  6 V ) 
 Front Panel Display 
2.989 V 
Vout – 9.5mV 
59.932 V 
Vout – 38mV 
 
2.994 V 
Vout – 4.5mV 
5.991 V 
Vout – 6mV 
_______V 
_______V 
_______V 
_______V 
 
_______V 
_______V 
_______V 
_______V 
3.011 V 
Vout + 9.5mV 
60.068 V 
Vout + 38mV 
 
3.006 V 
Vout + 4.5mV 
6.009 V 
Vout + 6mV 
Voltage Mode Regulation Test 
5V regulation with input current delta 1A to 60A 
 
 High Resistance Point     2 Ω   
 Low Resistance Point   0.50 Ω 
 
Resistance Range 2  
 Low Resistance Point  1.8 Ω 
 High Resistance Point 20 Ω  
 
Resistance Range 3  
 Low Resistance Point  18 Ω 
 High Resistance Point 200 Ω 
 
Resistance Range 4  
 Low Resistance Point    180 Ω 
 High Resistance Point   2000 Ω 
1.980 Ω 
0.486 Ω 
 
 
1.733 Ω 
19.360 Ω 
 
 
16.800 Ω 
159.88 Ω 
 
 
144 Ω 
1000 Ω 
______ Ω 
______ Ω 
 
 
______Ω 
______ Ω 
 
 
______Ω 
______Ω 
 
 
______Ω 
______Ω 
2.020 Ω 
0.514 Ω 
 
 
1.867 Ω 
20.640 Ω 
 
 
19.20 Ω 
240.12 Ω 
 
 
 298.8 Ω 
42,000 Ω