5 – Front Panel Operation
70
1.
Displays the minimum current
Displays the rms voltage
Displays the input power
Displays the maximum power
Displays the minimum power
Use and to scroll through the measurement selections
<chan> <value> A RMS
<chan> <value> WATTS
<chan> <value> W MAX
<chan> <value> W MIN
7 - Setting the GPIB Address
Your electronic load is shipped with the GPIB address set to 5. This address can only be changed from the front
panel using the Address menu located under the
Address key
Set the GPIB address as follows:
On the System keypad, press Address.
Enter the new address. For example, Press 7, Enter.
8 - Storing and Recalling Instrument States
You can save up to 10 states in memory and recall them from the front panel. All programmable settings are
saved. Only memory locations 0, 7, 8, or 9 are non-volatile. Data saved in locations 1 through 6 will be lost
when the unit is turned off.
NOTE: The SAV0 state is the power up state.
Save an instrument state in location 0 as follows:
Set the instrument to the state that you want to save.
Save this state to location 0. Press Shift, Save, 0, Enter.
Recall a saved state as follows:
Recall the state saved in location 0 by pressing Recall, 0, Enter.
Clear the non-volatile memory of the electronic load as follows:
1.
On the System keypad, press the Recall key and scroll to the reset command. Then
press Enter. This returns the unit to the factory-default settings.
*RST
Save these settings to location 0. Press Shift, Save, 0, Enter.
3. Repeat step #2 for memory locations 7 through 9. *SAV 7
*SAV 8