B - Performance Test and Calibration
84
Keysight N3302A Verification Test Record
Constant Current Mode Tests
30 Ampere Range Programming and Readback
Low Current ( 1 A )
Front Panel Display
High Current ( 30 A )
Front Panel Display
3 Ampere Range Programming and Readback
Low Current ( 1 A )
Front Panel Display
High Current ( 3 A )
0.989 A
Aout – 6.5mA
29.960 A
Aout – 21mA
0.994 A
Aout – 3.5mA
2.992 A
_______A
_______A
_______A
_______A
_______A
_______A
_______A
1.011 A
Aout + 6.5mA
30.040 A
Aout + 21mA
1.006 A
Aout + 3.5mA
3.008 A
Constant Current Regulation Test
2.5 Amp Regulation at input voltage delta 4V to 60V
60 Volt Range Programming and Readback
Low Voltage ( 3 V )
Front Panel Display
High Voltage ( 60 V )
Front Panel Display
6 Volt Range Programming and Readback
Low Voltage ( 3 V )
Front Panel Display
High Voltage ( 6 V )
Front Panel Display
2.989 V
Vout – 9.5mV
59.932 V
Vout – 38mV
2.994 V
Vout – 4.5mV
5.991 V
Vout – 6mV
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______V
3.011 V
Vout + 9.5mV
60.068 V
Vout + 38mV
3.006 V
Vout + 4.5mV
6.009 V
Vout + 6mV
Voltage Mode Regulation Test
5V regulation with input current delta 1A to 30A
High Resistance Point 4 Ω
Low Resistance Point 0.2 Ω
Resistance Range 2
Low Resistance Point 3.6 Ω
High Resistance Point 40 Ω
Resistance Range 3
Low Resistance Point 36 Ω
High Resistance Point 400 Ω
Resistance Range 4
Low Resistance Point 360 Ω
High Resistance Point 2000 Ω
3.944 Ω
0.1752 Ω
3.466 Ω
38.720 Ω
32.880 Ω
271.76 Ω
270 Ω
1200 Ω
______ Ω
______ Ω
______Ω
______Ω
______Ω
______ Ω
______Ω
______Ω
4.056 Ω
0.2248 Ω
3.734 Ω
41.280 Ω
39.120 Ω
528.24 Ω
720 Ω
10,000 Ω