Performance Test and Calibration - B
81
CR Mode Tests
These tests verify that the module operates in CR mode and the resistance programming is within specification.
The programmed resistance values are checked by recording the voltage across the current monitor resistor and
the input voltage (voltage at the modules + and – sense terminals ), then calculating the resistance as follows;
Load Resistance = Input Voltage / ( voltage across current monitor / monitor resistor value )
Resistance Range 1 Programming Accuracy
1. Connect Electronic Load, Power source and DVM’s as shown in figure B-4. Be sure + sense and sense are
connected to the +/- input terminals and the remote/local button is in the remote position.
2. Turn on the Electronic Load and press [ Recall ], scroll till display reads *RST and press [ Enter ].
3. Press [ Func ]. Scroll until display reads FUNC RES then press [ Enter ].
4. Press [ RES ]. Scroll until display reads RES:RANG, press [ 1 ] then [ Enter ].
5. Turn on the power source. See power source voltage and current values as listed in following table.
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
Power Source Voltage Setting
24V
48V
12V
30V
15V
30V
Power Source Current Setting
7A
7A
10A
10A
20A
7A
6. Wait 30 seconds. Checking resistance range 1, high resistance point. Calculate and record resistance on test card.
Input resistance = Input Voltage / ( DVM voltage reading from current shunt/ current shunt resistance).
7. Checking resistance range 1, low resistance point. Reset power source voltage to following values.
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
Power Source Voltage Setting
4V
6V
10V
5V
6V
5V
Power Source Current Setting
25A
10A
30A
50A
70A
25A
8. Press [ RES ] [ select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
0.2 Ω
1.2 Ω
0.5 Ω
0.125 Ω
0.100 Ω
0.250 Ω
9. Wait 30 seconds. Calculate and record resistance range 1, low resistance point.
Resistance Range 2 Programming Accuracy
1. Press [ RES ]. Scroll until display reads RES:RANG. For models N3302A, N3304A, N3305A, and N3306A press [ 9 ]
then [ Enter ]. For models N3303A and N3307A press [ 5 ] [ 0 ] then [ Enter ].
2. Checking resistance range 2, low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
3.6 Ω
44 Ω
1.8 Ω
4.5 Ω
0.9 Ω
9 Ω
3. Set power source voltage and current values as listed in following table.
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
Power Source Voltage Setting
20V
96V
6V
30V
6V
30V
Power Source Current Setting
7A
3A
7A
15A
8A
7A
4. Wait 15 seconds. Calculate and record resistance range 2 low resistance point.
5. Checking resistance range 2, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
40 Ω
480 Ω
20 Ω
50 Ω
10 Ω
100 Ω