B - Performance Test and Calibration
88
Keysight N3306A Verification Test Record
Constant Current Mode Tests
120 Ampere Range Programming and Readback
Low Current ( 1A )
Front Panel Display
High Current ( 120 A )
Front Panel Display
12 Ampere Range Programming and Readback
Low Current ( 1A )
Front Panel Display
High Current ( 12 A )
Front panel Display
0.9615 A
Aout – 20.5mA
119.8425 A
Aout – 80mA
0.9840A
Aout – 10.5mA
11.973 A
Aout – 16mA
_______A
_______A
_______A
_______A
_______A
_______A
_______A
_______A
1.0385 A
Aout + 20.5mA
120.1575 A
Aout + 80mA
1.0160 A
Aout + 10.5mA
12.027 A
Aout + 16mA
10 Amp Regulation at input voltage delta 4V to 60V
60 Volt Range Programming and Readback
Low Voltage ( 3 V )
Front Panel Display
High Voltage ( 60 V )
Front Panel Display
6 Volt Range Programming and Readback
Low Voltage ( 3 V )
Front Panel Display
High Voltage ( 6 V )
Front Panel Display
2.989 V
Vout – 9.5mV
59.932 V
Vout – 38mV
2.994 V
Vout – 4.5m
5.991 V
Vout – 6mV
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______V
3.011 V
Vout + 9.5mV
60.068 V
Vout + 38mV
3.006 V
Vout + 4.5m
6.009 V
Vout + 6mV
5V regulation with input current delta 1A to 120A
High Resistance Point 1 Ω
Low Resistance Point 0.100 Ω
Resistance Range 2
Low Resistance Point 0.9 Ω
High Resistance Point 10 Ω
Resistance Range 3
Low Resistance Point 9 Ω
High Resistance Point 100 Ω
Resistance Range 4
Low Resistance Point 90 Ω
High Resistance Point 1000 Ω
0.990 Ω
0.0936 Ω
0.8665 Ω
9.680 Ω
8.400 Ω
79.940 Ω
72 Ω
500 Ω
______ Ω
______ Ω
______Ω
______Ω
______Ω
______Ω
______Ω
______ Ω
1.010 Ω
0.1064 Ω
0.9335 Ω
10.32 Ω
9.60 Ω
120.06 Ω
149.4 Ω
21,000 Ω