Specifications - A
73
Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.
Table A-2. Supplemental Characteristics
N3302A
N3303A
N3304A
N3305A
N3306A
N3307A
Programming
Resolution
Constant Current Mode
Constant Voltage Mode
Constant Resistance
0.05mA/0.5mA
0.1mV/1mV
0.07/0.7/7/70m
0.02mA/0.2mA
0.4mV/4mV
0.82/8.2/82/
820m
0.1mA/1mA
0.1mV/1mV
0.035/0.35/3.5/
35m
0.1mA/1mA
0.25mV/2.5mV
0.085/0.85/8.5/
85m
0.2 mA / 2 mA
0.1mV / 1mV
0.0175/0.175/
1.75/17.5m
0.05mA/0.5mA
0.25mV/2.5mV
0.17/1.7/17/
170m
Readback Resolution
Current
Voltage
0.05mA/0.5mA
0.1mV/1mV
0.02mA/0.2mA
0.4mV/4mV
0.1mA/1mA
0.1mV/1mV
0.1mA/1mA
0.25mV/2.5mV
0.2 mA / 2 mA
0.1mV / 1mV
0.05mA/0.5mA
0.25mV/2.5mV
Slew Rates
1
Current Ranges
Slow band
Fast band 3V
Fast band <3V
500A/s - 25kA/s
50kA/s - 2.5MA/s
50kA/s - 250kA/s
167A/s - 8330A/s
16.7kA/s - 833kA/s
16.7kA/s - 83.3kA/s
1kA/s - 50kA/s
100kA/s - 5MA/s
100kA/s - 500kA/s
1kA/s - 50kA/s
100kA/s - 5MA/s
100kA/s - 500kA/s
2kA/s - 100kA/s
200kA/s - 10MA/s
200kA/s - 1MA/s
500A/s - 25kA/s
50kA/s - 2.5MA/s
50kA/s - 250kA/s
Voltage Ranges
Slow band
Fast band 3V
Fast band <3V
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
4kV/s - 200kV/s
400kV/s - 2MV/s
400kV/s - 200kV/
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
2.5kV/s - 125kV/s
250kV/s-1.25MV/s
250kV/s -125kV/s
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
2.5kV/s - 125kV/s
250kV/s-1.25MV/s
250kV/s -125kV/s
Resistance Range 1
Slow band
Fast band 3V
Fast band <3V
44/s - 1125/s
2250/s - 34k/s
2250/s - 3.4k/s
540/s - 13.5k/s
27k/s - 408k/s
27k/s - 40.8k/s
22/s - 560/s
1120/s - 17k/s
1120/s - 1.7k/s
55/s - 1400/s
2800/s -42.5k/s
2800/s -4.25k/s
11/s - 280/s
560/s - 8.5k/s
560/s - 850/s
110/s - 2800/s
5600/s - 85k/s
5600/s - 8.5k/s
Resistance Range 2
Slow band
Fast band 3V
Fast band <3V
440/s -11.25k/s
22.5k/s -340k/s
22.5k/s - 34k/s
5.4k/s - 135k/s
270k/s -4.08M/s
270k/s -408k/s
220/s - 5600/s
11.2k/s - 170k/s
11.2k/s - 17k/s
550/s - 14k/s
28k/s - 425k/s
28k/s - 42.5k/s
110/s - 2800/s
5600/s - 85k/s
5600/s - 8.5k/s
1.1k/s - 28k/s
56k/s - 850k/s
56k/s - 85k/s
Resistance Range 3
Slow band
Fast band 3V
Fast band <3V
4.4k/s-112.5k/s
225k/s - 3.4M/s
225k/s - 340k/s
54k/s - 1.35M/s
2.7M/s-40.8M/s
2.7M/s-4.08M/s
2.2k/s - 56k/s
112k/s - 1.7M /s
112k/s - 170k /s
5.5k/s - 140k/s
280k/s 4.25M/s
280k/s -425k/s
1.1k/s - 28k/s
56k/s - 850k/s
56k/s - 85k/s
11k/s - 280k/s
560k/s -8.5M/s
560k/s -850k/s
Resistance Range 4
Slow band
Fast band 3V
Fast band <3V
44k/s-1.125M/s
2.25M/s- 34M/s
2.25M
/s-3.4M
/s
540k/s -13.5M/s
27M/s - 408M/s
27M
/s -40.8M
/s
22k/s - 560k/s
1.12M/s -17M/s
1.12M
/s-1.7M
/s
55k/s - 1.4M/s
2.8M/s-42.5M/s
2.8M
/s-4.25M
/s
11k/s - 280k/s
560k/s -8.5M/s
560k
/s -850k
/s
110k/s -2.8M/s
5.6M/s - 85M/s
5.6M
/s -8.5M
/s
Slew Rate Accuracy
within 35% of programmed value
Programmable Short
66m max.
40m typical
200m max.
100m typical
33m max.
20m typical
33m max.
25m typical
17m max.
12m typical
33m max.
20m typical
Programmable Open
20k
80k
20k
80k
20k
80k
DC Isolation Voltage
±300 Vdc between + or - input binding post and chassis ground
Command Processing Time
Discrete commands
3 ms
List commands
1 ms
List Dwell Characteristics
Range
0 - 10 s
Resolution
1 ms
Accuracy
5 ms
1
Slew rate bands are not programmable. When you program a slew rate value outside the indicated bands, the electronic load will
automatically adjust the slew rate to fit within the band that is closest to the programmed value. The slew rate accuracy specification is
only applicable to slew rates within the indicated bands. Additionally,
when the transition from one setting to another is small, the small
signal bandwidth of the load limits the minimum transition time for all programmable slew rates (see chapter 2 for more information).
Below 3 volts, the maximum bandwidth of the electronic load is reduced by a factor of ten to one. For example, in the current range for
Model N3302A, the maximum slew rate is specified as 2.5MA/s, below 3 volts the maximum slew rate would be 250kA/s. Any slew rate
programmed between 2.5MA/s and 250kA/s would produce a slew rate of 250kA/s. Slew rates programmed slower than 250kA/s would
still correctly reflect their programmed value. If you are using transient mode to generate a high frequency pulse train, a reduced slew
rate might cause the load to never reach the upper programmed value before beginning the transition to the lower programmed value. So
even though the transient mode is still operational at lower voltages, a fast pulse train with large transitions may not be achievable.

Table of Contents

Related product manuals