B - Performance Test and Calibration
98
540 !
550 INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1
560 !
570 OUTPUT @Ld;"CAL:DATA";Vp1
580 !
590 OUTPUT @Ld;"CAL:LEV P2"
600 WAIT 1
610 !
620 INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2
630 !
640 OUTPUT @Ld;"CAL:DATA";Vp2
650 !
660 OUTPUT @Ld;"CAL:SAVE"
670 OUTPUT @Ld;"CAL:STAT OFF"
680 !
690 CLEAR SCREEN
700 PRINT "Voltage calibration adjustments complete"
710 PRINT "Verify Voltage Adjustments"
720 !
730 END