20 N5186A MXG and X-Series SA Measurement Guide
Basic Measurements
Making Measurements
Using the equivalent SCPI commands
On The MXG:
SYSTem:PRESet
RF1:FREQuency:CW 6GHZ
RF1:POWer:AMPLitude 0dBm
SIGNal:MODE AMODulation
SIGNal1:AMODulation:TYPE AM
SIGNal1:AM:SHAPe SINE
SIGNal1:AM:FREQuency 100KHZ
SIGNal1:AM 50
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the Signal Analyzer:
FREQuency:CENTer 6GHZ
FREQuency:SPAN 500KHZ
BANDwidth 100 Hz
CALCulate:MARKer1:MODE DELTa
CALCulate:MARKer1:MAXimum:RIGHt
Repeat the above command until the marker is at the next highest peak.
To retrieve the delta marker:
CALCulate:MARKer1:Y?