38 N5186A MXG and X-Series SA Measurement Guide
Basic Measurements
Making Measurements
Using the equivalent SCPI commands
On the MXG:
SYSTem:PRESet
RF1:FREQuency:CW 6GHZ
RF1:POWer:AMPLitude 0dBm
SIGNal1:MODE WAVeform
SIGN1:WAV "/Examples/GSM_1C_Burst.wfm"
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the Signal Analyzer:
FREQuency:CENTer 6GHZ
FREQuency:SPAN 900KHZ
ACPower:BANDwidth 470 Hz
DISPlay:TXPower:WINDow1:TRACe:MAXHold ON
On the MXG:
SIGN1:WAV "/Examples/WCDMA_TM1_64DPCH_4C.wfm"
SIGNal1 ON
RF1:OUTPut:STATe ON
On the Signal Analyzer:
ACPower:BANDwidth 620 Hz
FREQuency:SPAN 30MHZ