Keysight NFA Series Noise Figure Analyzers Service Guide 129
Instrument Messages
Condition Messages
Condition Errors 701 to 799, Operation
This series of errors corresponds to the bits in the STATus:OPERation
register, which can be read with a STATus:OPERation? event query or a
STATus:OPERation:CONDition? query.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated.
For example, error 721 indicates that the DC Coupled condition has been
detected; error 1721 indicates that condition has been cleared.
Unless otherwise noted in the Error or Warning column below, these are status
bits only, with no corresponding error message or number
619 9 Integrity status bit
only
This bit is the summary bit for the
STATus:QUEStionable:INTegrity sub-register.
621 10 unused
623 11 unused
625 12 unused
627 13 unused
629 14 unused
Err# Bit in
status
register
Message Error or
Warning
More Information
Err# Bit in
status
register
Message Error or
Warning
More Information
701 0 Calibrating status bit
only
703 1 Settling status bit
only
705 2 unused
707 3 Sweeping status bit
only
709 4 Measuring status bit
only
711 5 Waiting for Trigger status bit
only
713 6 Waiting for Arm status bit
only