Keysight NFA Series Noise Figure Analyzers Service Guide 87
Instrument Messages
Event Messages
–221 Settings conflict;
Destination trace for
Trace Math cannot be
a trace operand
The resulting trace data (from doing a trace math function) cannot be put into
the any of the traces that are being used by the math operation
–221 Settings conflict;
Downconv only
available when DUT is
Amplifier
SCPI only message. The System Downconverter can only be set to ‘On’ when
the DUT type is amplifier. Change DUT type to Amplifier if the System
Downconverter is required
–221 Settings conflict;
EDGE EVM only
supports EDGE TCH
burst type.
–221 Settings conflict;
EMI Detectors are not
available in FFT
sweep
QPD, EMI Average, EMI Peak, and MIL Peak are not allowed when in the
manually selected FFT sweep mode
–221 Settings conflict;
External Mixer not
available
A command has been sent to reference the External Mixer in a model that
does not contain it
–221 Settings Conflict;
FAST method can only
be used while Radio
Std is W-CDMA
–221 Settings conflict;
Feature not available
in this View
Some functionality is available in one View, but not in another. (See the Views
under the
Mode/Measurement/View selection.) This error occurs if
you send a SCPI command or push a grayed-out key that is not available in
the current selected View
–221 Settings conflict;
Feature not supported
for selected source
You have selected a feature that the selected source does not support
–221 Settings conflict;
Feature not supported
for this Input.
Some functionality is not available when certain Inputs are selected. For
example, Trigger Holdoff is not available for the BBIQ input
–221 Settings conflict;
Feature not supported
for this measurement.
Some functionality is available in one measurement, but not in another. (See
the measurements under the
Mode/Measurement/View selection.)
This error occurs if you send a SCPI command or push a grayed-out key that
is not available in the current selected measurement
–221 Settings conflict;
Feature not supported
for this model number
This functionality is not a part of the instrument you are using, but may be
found in other models in the X-Series
Err# Message Verbose/Correction Information