126 Keysight N9010B EXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
309 4 User Cal; Adjusted 
for new RBW
W The measurement RBW has been changed since 
the last calibration (~CAL)
311 5 Calibration W
311 5 Calibration; ENR 
table extrapolated
W One or more calibration or measurement 
frequency points exceed the currently loaded Cal 
or Meas ENR Table frequency ranges. The 
corresponding ENR table’s lowest frequency ENR 
value will be re-used for frequencies less than the 
table range, and the highest frequency ENR value 
will be re-used for frequencies greater than the 
table range.
(~ENR)
311 5 Calibration; No ENR 
data present
W No ENR Data (ENR)
313 6 Source Uncal W
313 6 Source Uncal;adj 
Start Freq or RBW
W While using a Tracking Source, you must make 
sure the Start Frequency is high enough to avoid 
capturing LO feedthrough in the trace. This 
depends on both Start Freq and RBW. If you get 
this message, increase the Start Freq or narrow 
the RBW.
315 7 Preamp: Accy unspec’d 
<XX kHz
W Preamp will function at lower frequencies, but the 
performance is not specified below XX kHz (XX is 
model number specific).
317 8 unused W
319 9 unused W
321 10 unused W
323 11 unused W
325 12 unused W
327 13 unused W
329 14 unused W
Err# Bit in 
status 
register
Message Error or 
Warning
More Information