EasyManua.ls Logo

Keysight N9010B - Displayed Average Noise Level (DANL)

Keysight N9010B
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
566 Keysight N9010B EXA Signal Analyzer Service Guide
Functional Tests
Displayed Average Noise Level (DANL)
Displayed Average Noise Level (DANL)
Test Limits (with 0 dB input attenuation)
See Table 17-2 for values.
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50 Ω load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Figure 17-1 DANL Test Setup
Procedure
1. Configure the equipment as shown in Figure 17-1
2. Press MODE/MEAS. Select Spectrum Analyzer mode, Swept SA
measurement, Normal view. Press OK. Press Mode Preset on the analyzer.
3. Set up the signal analyzer by pressing:
FREQ, Center Frequency, 10, MHz
Input/Output, Input, RF Coupling, select DC
FREQ, Span, 10, kHz
AMPTD, –70, dBm
AMPTD, Attenuation, Mech Atten, 0, dB
Table 17-2 DANL Measurement Requirements
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Termination, 50
Ω Type-N (m)
a
Frequency: 10 MHz to 18 GHz 909A Option 012
Termination, 50
Ω 2.4 mm (f)
b
Frequency: 10 MHz to 50 GHz 85138B
a. For Options 503, 507, 513, 526
b. For Options 532, 544