Keysight N9010B EXA Signal Analyzer Service Guide 67
Boot Up and Initialization Troubleshooting
Signal Level Verification
Signal Level Problem with
Input Frequencies > 3.6 GHz
Measure the 4.8 GHz RF calibrator signal level by pressing Input/Output, RF
Calibrator, 4.8 GHz. Now press FREQ, 4.8 GHz, Span, 1 MHz, Peak Search. If
the analyzer is functioning correctly in high band, the 4.8 GHz calibrator level
should be 28 dBm ± 2 dB. See the figure below.
Is the signal level correct?
If yes:
The analyzer high band path is functioning properly at least at 4.8 GHz.
If no:
Refer to Chapter 4, “RF Section Troubleshooting (RF/Microwave
Analyzers)” in this manual.
It is possible that other input frequencies > 3.6 GHz could have an
amplitude problem even though the 4.8 GHz calibrator is within tolerance.
Using the internal RF calibrator provides a quick check of the high band
path.