146 Keysight N9010B EXA Signal Analyzer Service Guide
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Troubleshooting a Low Band Problem
1. Reference Assembly Verification
Remove cable W19 from A9 Input Attenuator A (1) and measure the 50 MHz
calibrator signal on the cable end with a functioning Spectrum Analyzer. Refer
to Figure 4-5.
Figure 4-5 W19 Location