Keysight N9010B EXA Signal Analyzer Service Guide 505
Assembly Replacement Procedures
Power Supply Assembly
Power Supply Assembly
Removal
1. Remove the instrument outer case. Refer to the “Instrument Outer Case”
removal procedure.
2. Remove the rear panel. Refer to the “Rear Panel” removal procedure.
3. Remove the top brace and reference bracket. Refer to the “Top Brace”
removal procedure.
4. Remove the reference assembly. Refer to the “Reference Assembly”
removal procedure.
5. Remove the CPU assembly. Refer to the “CPU Assembly” removal
procedure.
6. Refer to Figure 15-55. Remove the three screws (1) attaching the power
supply (2) to the power supply bracket (3).
7. The power supply assembly can be removed from the chassis by pulling
straight out the back.
Figure 15-55 Power Supply Assembly Removal