Keysight N9010B EXA Signal Analyzer Service Guide 457
Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
Low Band Switch and Coaxial Switches - RF/Microwave Instruments
with Options DP2, MPB, B40
Removal
1. Refer to Figure 15-12. Remove semi-rigid coax cables W32 and W33.
Figure 15-12 Remove W33 and W32