Keysight N9010B EXA Signal Analyzer Service Guide 565
Functional Tests
Test Equipment
P03 Low Power Sensor Frequency: 50 MHz to 3.6 GHz
Amplitude: –70 to –10 dBm
8481D 8485D
8487D
Pxx other
than P03
Low Power Sensor
Frequency: 50 MHz to 44 GHz
c
Amplitude: –70 to –10 dBm
8487D
Miscellaneous Equipment
All Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
0955-0306
526 Power Splitter, 3.5 mm Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667B
532, 544 Power Splitter Nominal Insertion Loss: 8.5 dB
Tracking Between Ports: < 0.4 dB
11667C
503, 508,
513
Power Splitter Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667A 11667B
All Termination, 50? Type N (m) Connector
Frequency: 30 Hz to 26.5 GHz
909A Option 012
532, 544 Termination, 50? 2.4 mm (f) Connector
Frequency: 9 kHz to 44 GHz
85138B
a. Maximum frequency equal to or greater than the maximum frequency of the EXA.
b. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540, E8257D
Option 520, E8257D Option 550, E8257D Option 567, E8267D Option 544.
c. Maximum frequency equal to or greater than the maximum frequency of the preamplifier option.
Table 17-1
Analyzer
Option
Item Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model