Keysight N9010B EXA Signal Analyzer Service Guide 153
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Turn the instrument on and allow it to complete its full boot up process to
the signal analyzer application.
Turn Auto Align off by pressing System, Alignments, Auto Align, Auto
Align, Off on the analyzer.
Verify all power supply LED's on the Front End Troubleshooting board are
turned on. Under the J2 connector, the supplies are +15 VDC, +9 VDC, +5.2
VDC, and 15 VDC. Under the J1 connector, the supplies are +15 VDC, 15
VDC, +9 VDC and +5.2 VDC. Verify both input attenuator power supplies.
DS9 under J3 should be on if the +25V supply is up for Attenuator A. DS10
under J4 should be on if the +25V supply is up for Attenuator B. If any of the
power supply LED's are not turned on, see Chapter 2, “Boot Up and
Initialization Troubleshooting”.
4. Input Attenuator A Control Logic Verification
—Press AMPTD, Attenuation, Mech Atten 0 dB
All the attenuation LED's on the Front End Troubleshooting board should be off
except for DS9 and DS10, +25V supply for Attenuator A and Attenuator B.
Refer to Figure 4-12.
Resistors on the board can get very hot.
The analyzer will display an error dialogue box when booting up. Simply
click “OK” so the instrument can complete its boot process. Additionally,
the analyzer noise floor will be near the top of the display since the RF
assemblies are not powered up. This is normal with the Front End
Troubleshooting board connected to the A15 Front End Driver Assembly