216                                Keysight N9010B EXA Signal Analyzer Service Guide
RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
The Second LO input was tested during the Second LO Verification.
The LO input at W4, and the LO outputs were tested when performing the First 
LO Verification.
It is possible to measure the A13W1 input cable to A13A1 where it connects to 
A13A1J2, and the output of the A13A1 at A13W2 where it connects to A13J11. 
A13A1 IF Band Pass Filter is not separately replaceable, but being able to 
determine where the signal level is faulty, can aid in verifying the faulty 
assembly. See the block diagram for signal levels.
Electronic Attenuator Test (Option EA3)
The electronic attenuator is aligned as part of the System Gain internal 
alignment process. See the description of the initial alignments and the 
location of the alignment history file in the Boot Up and Initialization 
Troubleshooting chapter. Viewing the Alignment History file will tell you if 
other alignment tests failed, and reveal which electronic attenuator steps 
failed. 
To verify the electronic attenuator operation on instruments with Option EA3, 
set up the analyzer to view the 50 MHz calibrator signal on screen. Press 
Input/Output, RF Calibrator, 50 MHz, FREQ, Center Frequency, 50 MHz, 
Span, 500 kHz
Press Peak Search to place a marker on the calibrator signal, then press 
Marker Delta 
Activate the electronic attenuator. Press AMPTD, Attenuation, Elec Atten, 
Enabled.
Press the up arrow key to step through all attenuator steps while monitoring 
the marker delta readout. The signal level should not change more than 0.4 dB 
as indicated by the marker delta readout. 
If the attenuator switches slowly, or the signal is not present when the 
electronic attenuator is switched into the signal path or the signal appears only 
when the attenuator is switched on, suspect the control signals from the A15 
Front End Controller are faulty or the A13 Front End assembly is faulty. If the 
attenuator switches, but the switching error between attenuator settings is 
greater than ± 0.4 dB, assure that any other internal alignment failures are 
resolved since the System Gain alignment must run before optimum 
performance is possible. Assure you have evaluated all causes before you 
replace the Front End assembly.
If you determine the problem is at a frequency other than 50 MHz, perform the 
low band flatness adjustment before replacing the Front End assembly.