Keysight N9010B EXA Signal Analyzer Service Guide 29
Overview
Service Equipment You Will Need
High Sensitivity
Millimeter Power Sensor
(for Options 532)
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: 70 to 20 dBm
Zero Set: ± 20 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 4.5%
VSWR:
50 MHz to 100 MHz: 1.19:1
100 MHz to 4 GHz: 1.15:1
4 GHz to 12.4 GHz: 1.20:1
12.4 GHz to 18 GHz: 1.29:1
18 GHz to 34 GHz: 1.37:1
34 GHz to 40 GHz: 1.61:1
40 GHz to 50 GHz: 1.89:1
Input Connector: 2.4 mm (m)
8487D A, P
Low Power Microwave
Power Sensor
(2 required)
(for Option P07, P13,
or P26 only)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: 70 to +20 dBm
Zero Set: ± 50 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
8485D
e
A, P
Power Sensor Cable
(2 required)
11730A A, P, T
Standards
Frequency Standard Frequency: 10 MHz
Accuracy: < ±1 e10
10
Microsemi
5071A-C002
Keysight 5061B,
5071A
Symmetricom
5071A-C001
A, P
50 MHz,
25 dBm Calibrator
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: 45 dBc
Z5602B
(Option H51)
Z5602A
(Option H51)
A
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Model
a
Alternative
Model
Use
b