Keysight N9010B EXA Signal Analyzer Service Guide 553
Post-Repair Procedures
Post-Repair Procedures
Data Restore
1. Connect a USB mouse to one of the instrument front panel USB ports.
2. Locate the USB FLASH drive that was earlier used to backup the
instrument calibration file and insert it into one of the instrument front
panel USB ports.
3. Restore the instrument calibration data by pressing: System, Alignments,
Backup or Restore Align Data... The Alignment Data Wizard window as
shown in Figure 16-1 will appear.
4. Follow the on-screen instruction to restore the previously saved
calibration data from the USB FLASH drive.
5. Cycle instrument power.