Keysight N9010B EXA Signal Analyzer Service Guide 95
Instrument Messages
Event Messages
–221 Settings conflict;
System Display
Settings, Annotation
is Off
This is an override that turns off many of the annotations. This is available as a
security feature
–221 Settings conflict; T
hot must be greater
than T cold
The Tcold value set under
Meas Setup, ENR, Tcold needs to be lower
than the Thot value currently being set. Tcold is often taken as the ambient
temperature of the noise source. If using an SNS the Tcold value may be read
automatically before every sweep
–221 Settings Conflict;
The parameter cannot
be changed in FAST
mode
–221 Settings conflict;
Trace Math is not
available while
Normalize is on
The Normalize function works by doing trace manipulation, so trace math is
not available while normalization is running
–221 Settings conflict;
Tracking Source
unavailable in FFT
Sweeps
Since FFTs do not sweep, you cannot use a Tracking Source while doing FFTs
–221 Settings conflict;
Trigger input in use
for source
synchronization
If Point Trigger is being used with an external trigger input to synchronize an
external source to the analyzer, that trigger input is unavailable for triggering
–221 Settings conflict;
Trigger is not
available with span >
0 Hz.
–221 Settings conflict; Tx
Band Spur meas does
not support this
frequency band.
The transmit band spur measurement does not support all of the
commercially available frequency bands. You need to change your selection
under
Mode Setup, Radio, Band to one of the supported bands
–221 Settings conflict; Tx
Band Spur measurement
is not defined for
mobiles.
Only base station testing is available
–221 Settings
conflict;Administrat
or privileges
required
You must be logged in with administrator privileges to do this. Log out and
log back in as the Administrator, then restart the SA application