Instrument Messages
Event Messages
–200 to –299, Execution Errors
This section lists all messages in the range except –221 error messages. For
details of –221 messages, see “–221 Settings Conflict Errors” on page 89.
Note that Execution Errors are divided into subclasses, as follows:
–221 Settings
conflict;Range <0> is
turned off as total
range points > 40001
Max of Total range points is 400000. Reduce Scan Points or increase Step
Size in order to turn on that range
–221 Settings conflict;RF
Input 2 is not
available above 1GHz
If the frequency range is set above 1GHz, you cannot change to RF Input 2
–221 Settings
conflict;Scan Time &
Points do not apply
in current Step/Time
Control.
The Scan Time & Points are not available when Step/Time Control is set to
Step & Dwell
–221 Settings
conflict;Step Size &
Meas Time do not
apply in current
Step/Time Control.
The Step Size & Meas Time are not available when Step/Time Control is set to
Scan Time&Pts
Err# Message Verbose/Correction Information
–21x Trigger errors
–22x Parameter error
–23x Data corrupt or stale (invalid data)
–24x Hardware error
–25x Mass storage error
–26x Expression data error
–27x Macro error
–28x Program error (a downloaded program-related execution error)
–29x Memory use error
Err# Message Verbose/Correction Information
–200 All ranges are off.
Turn on at least a
range
There are no range turn on in scan table. You need to turn on at least a range to
initiate a scan
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom

Table of Contents