Instrument Messages
Condition Messages
Condition Errors 701 to 799, Operation
This series of errors corresponds to the bits in the STATus:OPERation 
register, which can be read with a STATus:OPERation? event query or a 
STATus:OPERation:CONDition? query.
An event with the error number shown in the table means the condition has 
been detected. When the condition is cleared, an event with the error number 
plus 1000 is generated.
For example, error 721 indicates that the DC Coupled condition has been 
detected; error 1721 indicates that condition has been cleared.
Unless otherwise noted in the Error or Warning column below, these are status 
bits only, with no corresponding error message or number
619 9 Integrity status bit 
only
This bit is the summary bit for the 
STATus:QUEStionable:INTegrity sub-register.
621 10 unused 
623 11 unused
625 12 unused 
627 13 unused
629 14 unused
Err# Bit in 
status 
register
Message Error or 
Warning
More Information
Err# Bit in 
status 
register
Message Error or 
Warning
More Information
701 0 Calibrating status bit 
only
703 1 Settling status bit 
only
705 2 unused
707 3 Sweeping status bit 
only
709 4 Measuring status bit 
only
711 5 Waiting for Trigger status bit 
only
713 6 Waiting for Arm status bit 
only