22
Overview
ESD Information
Handling of Electronic Components and ESD
The possibility of unseen damage caused by ESD is present whenever
components are transported, stored, or used. The risk of ESD damage can be
greatly reduced by paying close attention to how all components are handled.
Perform work on all components at a static-safe workstation.
Keep static-generating materials at least one meter away from all
components.
Store or transport components in static-shielding containers.
CAUTION Always handle printed circuit board assemblies by the edges. This will reduce
the possibility of ESD damage to components and prevent contamination of
exposed plating.
Test Equipment Usage and ESD
Before connecting any coaxial cable to an analyzer connector, momentarily
short the center and outer conductors of the cable together.
Personnel should be grounded with a 1 megohm resistor-isolated
wrist-strap before touching the center pin of any connector and before
removing any assembly from the analyzer.
Be sure that all analyzers are properly earth-grounded to prevent build-up
of static charge.
For Additional Information about ESD
For more information about preventing ESD damage, contact the Electrical
Over Stress/Electrostatic Discharge (EOS/ESD) Association, Inc. The ESD
standards developed by this agency are sanctioned by the American National
Standards Institute (ANSI).