Assembly Replacement Procedures
Power Supply Assembly
Power Supply Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer
Caseremoval procedure.
2. Remove the rear panel. Refer to the Rear Panelremoval procedure.
3. Remove the top brace and reference bracket. Refer to the Top Brace and
Reference Bracket removal procedure.
4. Remove the reference assembly. Refer to the Reference Assembly removal
procedure.
5. Remove the CPU assembly. Refer to the CPU Assembly removal
procedure.
6. Refer to Figure 14-46. Remove the three screws (1) attaching the power
supply (2) to the power supply bracket (3).
7. The power supply assembly can be removed from the chassis by pulling
straight out the back.