EasyManua.ls Logo

Keysight N9020A - Replacement

Keysight N9020A
642 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
493
Assembly Replacement Procedures
Power Supply Assembly
Figure 14-46 Power Supply Assembly Removal
Replacement
1. Slide the power supply assembly into the slot at the rear of the instrument
and push on the assembly to mate the connectors to the midplane
assembly.
2. Refer to Figure 14-46. Replace the three screws (1) through the power
supply bracket (3) and into the power supply (2). Torque to 9
inch-pounds.
3. Replace the CPU assembly. Refer to the CPU Assembly replacement
procedure.
4. Replace the reference assembly. Refer to the Reference Assembly
replacement procedure.
5. Replace the top brace and reference bracket. Refer to the Top Brace and
Reference Bracket replacement procedure.
6. Replace the rear panel. Refer to the Rear Panelreplacement procedure.