Functional Tests
Scale Fidelity
Figure 16-7 Scale Fidelity Setup
NOTE Averaging is used for all measurements to improve repeatability and reduce
measurement uncertainty.
Procedure
1. Configure the equipment as shown in Figure 16-7.
2. Preset the Source and press Mode, Spectrum Analyzer, Mode Preset on
the analyzer.
3. Set up the synthesized sweeper by pressing:
Frequency, 50, MHz
Amplitude, +5, dBm
RF On/Off, On
4. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 1, MHz
AMPTD Y Scale, Ref Level, 0, dBm
Meas Setup, Average/Hold Number, 10, Enter
Trace/Detector, Trace Average
Peak Search
5. Set the external 10 dB step attenuator to 0 dB.