CPU/Disk Drive Troubleshooting
Reloading the X-Series Software
Reloading the X-Series Software
The X-Series software contains all the required components for the signal 
analyzer application as well as all software options. If the X-Series software has 
become corrupt the Windows operating system will boot but the X-Series 
software application will fail to start. In these cases go to Chapter 17, 
“Instrument Software.” for information on how to update the X-Series 
software. This procedure will not affect the instrument’s calibration or user 
files.
If this does not resolve the boot issue, or the instrument never boots the 
Windows operating system proceed to the “Disk Drive Recovery Process”.